Illinois Bids > Bid Detail

PtychoProbe XRF Detector Positioner

Agency: ENERGY, DEPARTMENT OF
Level of Government: Federal
Category:
  • 66 - Instruments and Laboratory Equipment
Opps ID: NBD00159495217390974
Posted Date: Apr 13, 2024
Due Date: Apr 18, 2024
Solicitation No: 4-B138-Q-00398-00
Source: https://sam.gov/opp/903202a3b7...
Follow
PtychoProbe XRF Detector Positioner
Active
Contract Opportunity
Notice ID
4-B138-Q-00398-00
Related Notice
Department/Ind. Agency
ENERGY, DEPARTMENT OF
Sub-tier
ENERGY, DEPARTMENT OF
Office
ARGONNE NATL LAB - DOE CONTRACTOR
Looking for contract opportunity help?

APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.

APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.

The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).

General Information
  • Contract Opportunity Type: Solicitation (Original)
  • All Dates/Times are: (UTC-05:00) CENTRAL STANDARD TIME, CHICAGO, USA
  • Original Published Date: Apr 12, 2024 08:29 pm CDT
  • Original Date Offers Due: Apr 18, 2024 05:00 pm CDT
  • Inactive Policy: Manual
  • Original Inactive Date: Apr 20, 2024
  • Initiative:
    • None
Classification
  • Original Set Aside:
  • Product Service Code: 6640 - LABORATORY EQUIPMENT AND SUPPLIES
  • NAICS Code:
    • 423490 - Other Professional Equipment and Supplies Merchant Wholesalers
  • Place of Performance:
    Lemont , IL 60439
    USA
Description

This solicitation is for a XYZ (three axis) manipulator to position a x-ray detector in a vacuum chamber (1E-7 mbar). The device will bolt to a fixed flange on the side of the vacuum chamber and the moving flange to an x-ray detector. The following specifications are to be met:




  • X and Y travel of 25 mm with 5 micron or better resolution

  • Z travel of 400 mm with 5 micron or better resolution

  • Clear aperture shall allow for passage of a 38 mm diameter sensor head

  • The fixed flange is a 6” OD CF and the moving flange is 4.5” OD CF

  • All axes are stepper motor driven

  • All axes must have absolute encoding

  • All axes must be guided by profile rail type guides, no ball bushings are allowed

  • The moving flange load capacity should be 137 N in the vertical and 14 N*m about the horizontal axis




Attachments/Links
Contact Information
Contracting Office Address
  • 9700 S. Cass Ave
  • Lemont , IL 60439
  • USA
Primary Point of Contact
Secondary Point of Contact


History
  • Apr 12, 2024 08:29 pm CDTSolicitation (Original)

TRY FOR FREE

Not a USAOPPS Member Yet?

Get unlimited access to thousands of active local, state and federal government bids and awards in All 50 States.

Start Free Trial Today >